SCOPE ACCREDITED CALIBRATION LABORATORY
PT. Tri Artha Manunggal
TEMPERATURE
No
|
Instrument to be
Calibrated |
Measurement
Range |
Best
Measurement Capability *) |
Methods /
Specification |
Notes
|
1
|
Temperature Sensor
With display unit |
0° C - 300° C
|
0.6° C
|
|
|
|
|
300° C - 600° C
|
1° C
|
JIS 8710-1993
|
|
2
|
Temperature Enclosure,
Oven |
0° C - 350° C
|
0.84° C
|
KAN Pd-02.04
|
|
|
Furnace
|
400° C - 1000° C
|
1.8° C
|
|
|
|
Waterbath
|
0° C - 95° C
|
1.1° C
|
|
|
3
|
Termocouple + Indicator.
|
20° C - 100° C |
0,24° C - 0,67° C
|
PK-T-02
|
|
|
Digital Thermometer
|
100° C - 400° C
|
|
||
|
Temperature Gauge
|
400° C - 950° C
|
3,0° C
|
|
|
4
|
Infrared Thermometer |
50° C - 500° C
|
3,5° C |
PK-T-04
|
|
5 | Thermohygrometer Suhu | 15° C - 40° C | 0.59° C | PK-T-03 |
|
Kelembapan | 25% - 90% | 2,5% of reading | |||
Notes :
Best Measurement Capability is stated as Expanded Ucertainty at Approximately 95% confident level, calculated using Coverage factor of K = 2, includes Ucertainty contribution due to reference standards used by the laboratory, measurement prosess at the laboratory, and contribution due to typical characteristic of nearly ideal device under calibration respectif to the laboratory.
The actual characteristic of device under calibration shall also be considered for individual calibration certificate issued by to the laboratory.
DIRECTOR :
Mr. I Ketut Suada
Laboratory :
Panorama Serpong Blok C2 No. 12-12A, Baktijaya, Setu, Tangerang Selatan 15315
Telephone :
021-2930 8235
Website :
http//www.tri-artha.com
Email :
info@tri-artha.com
Representative Office :
Jl. PJHI Batakan No. 70A RT 15, Manggar, Kec. Balikpapan Timur, Kota Balikpapan, Kalimantan Timur 76116
BANKING :
BANK MANDIRI
BANK BCA